Ambul ventilator 6000S
This product is no longer available.
Main unit, ventilation hose, patient valve-mask
Portable and easy operation
Innovative voice-guided direction
Tidal volume range: 100- 1500ml
This product is not Approved by Health Canada
Portable Invasive / Non-invasive ventilator
Suitable for vehicle transport, long-distance transport, hospital transport, and wilderness rescue.
TFT display of airway pressure and ventilation mode
3 ventilation modes. IPPV, Assisted control and Manual control
Li-ion rechargeable battery with more than 10 hours of working time
Air Mix option for 60% 02
3 knobs for flexible parameters adjustment
Device category | Class Ilb |
Degree of Protection against Electric Shock | BF |
Degree of protection against Ingress of Liquids | IPX4 |
Degree of Protection against Hazards of Explosion | Ordinary equipment, without protection against explosion; not for use with flammable anesthetic |
Mode of Operation: | Continuous running equipment |
Dimensions WxHxD in mm | 240x 120x 100mm include connectors |
Weight | Approx. 1.3kg |
Display | 2.4" TFT color display resolution ratio: 320.'240 |
Pressurized gas connection | External thread 9/16-18 |
Ventilation hose connection | External diameter 15mm/Internal diameter 20mm |
Operating: Temperature range Humidity Air pressure | -20'C to +50°C 15% to 95% 70 kPa to 110 kPa |
Store/transport: Temperature range Humidity Air pressure | --40°C to +50°C 595% 70 kPa to 110 kPa |
Power supply | AC: 100 to 240V: 1.5 to 4A: 50/ 60HZ DC : 12 V Adapter |
Work current | Imin=0.3A; Irn..0.6A |
Battery: Standard Work temperature | Lithium battery 7.4V: 3400mAh : operating time: more than 10 hour -20°C to +55°C |
Ventilation modes | IPPV: A/C: CPR |
Operating pressure | 2.7 to 6.0 bar |
Required gas supply Standard gas supply: -Supply pressure -Drawn flow Optimal gas supply: -Supply pressure -Drawn flow Non-recommended gas supply: -Supply pressure -Drawn flow | At least 2.7bar At least 70 1/min Oxygen(ATPD) At least 4.5bar At least 100 1/min Oxygen(ATPD) Less than 2.7bar Less than 80 l/min Oxygen(ATPD) |
Insp-exp. Ratio | Constant 1:1.67 |
Minute volume(MV) | Continuously variable from 3 to 20 Umin(ATPD) |
Ventilation frequency | Continuously variable from 5to 40 min' |
02 concentration | No Air Mix 100% and Air Mix (see page 42) |
MV tolerances | ±20% |
Max. ventilation pressure | 20 to 60 mbar |
Safety airway pressure | 575 mbar |
Pressure gauge accuracy | 0 to 60 mbar Deviation ±5% |
Trigger sensitivity | -2 mbar Deviation ±0.5mbar |
Reusable ventilation hose | Spiral silicone |
Patient valve resistance: Inspiration Expiration Emergency air intake | <6 mbar at 30, 60 1/min <6 mbar at 30, 60 1/min <6 mbar at 15. 30 1/min |
Respiratory compliance | 100 ml/cmH2O |
Voice alarm mute time | 120s |
Voice prompt language | Support five languages, default Simplified Chinese |